288
Chapter 5
Language Reference
SENSe Subsystem
Adjacent Channel Power—Automatic Sweep Time
[:SENSe]:ACP:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:ACP:SWEep:TIME:AUTO?
Sets the sweep time to be automatically coupled for the fastest
measurement time.
Factory Preset
and *RST:
On
Remarks:
You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
History:
Revision A.03.00 or later, in cdmaOne revision A.04.00
Adjacent Channel Power—Trigger Source
[:SENSe]:ACP:TRIGger:SOURce
EXTernal[1]|EXTernal2|FRAMe|IF|IMMediate|RFBurst
[:SENSe]:ACP:TRIGger:SOURce?
Select the trigger source used to control the data acquisitions.
External 1 – front panel external trigger input
External 2 – rear panel external trigger input
Frame – internal frame trigger from front panel input
IF – internal IF envelope (video) trigger
Immediate – the next data acquisition is immediately taken,
capturing the signal asynchronously (also called free run).
RF Burst – wideband RF burst envelope trigger that has automatic
level control for periodic burst signals.
Factory Preset
and *RST:
Immediate for BS
RF Burst for MS
Remarks:
You must be in Basic, cdmaOne, iDEN, NADC, or PDC
mode to use this command. Use INSTrument:SELect to
set the mode.
In Basic mode, for offset frequencies >12.5 MHz, the
external triggers will be a more reliable trigger source
than RF burst. Also, you can use the Waveform
measurement to set up trigger delay.
Summary of Contents for E4406A VSA Series
Page 4: ...4 ...
Page 59: ...59 2 Programming Fundamentals ...
Page 124: ...124 Chapter2 Programming Fundamentals Using the LAN to Control the Analyzer ...
Page 125: ...125 3 Programming Examples ...
Page 164: ...164 Chapter3 Programming Examples Using Java Programming Over Socket LAN ...
Page 165: ...165 4 Programming Command Cross References ...
Page 379: ...379 6 Error Messages ...
Page 412: ...412 Chapter6 Error Messages Error Message Descriptions ...