Chapter 5
305
Language Reference
SENSe Subsystem
Correction for Mobile Station RF Port External
Attenuation
[:SENSe]:CORRection:MS[:RF]:LOSS <rel_power>
[:SENSe]:CORRection:MS[:RF]:LOSS?
Set the correction equal to the external attenuation used when
measuring mobile stations.
Factory Preset
and *RST:
0.0 dB
Range:
−
50.0 to 50.0 dB
Default Unit:
dB
Remarks:
You must be in the cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & Arib), iDEN, NADC or PDC mode to
use this command. Use INSTrument:SELect to set the
mode.
Value is global to the current mode.
Select the Input Port
[:SENSe]:FEED IONLy|IQ|RF|IFALign|AREFerence
[:SENSe]:FEED?
Select the input port.
IONLy is the Iin-phase component of an IQ signal
IQ is the IQ Input port
RF in the RF INPUT port
IF Align is the IF alignment signal source (internal, 321.4 MHz)
Amplitude Reference is the internal amplitude reference source
(50 MHz)
Factory Preset
and *RST:
RF
Front Panel
Access:
Input, Input Port