Specifications
1
E4416A/E4417A Power Meters Service Guide
5
Video Bandwidth/ Dynamic Range Optimization
The power measurement system- comprising the sensor and meter- has its
maximum video bandwidth defined by the E- Series E9320 power sensor. To
optimize the system's dynamic range for peak power measurements, the Video
Bandwidth in the meter can be set to
High
,
Medium
, or
Low
as detailed in
Table 1- 1
. The filter video bandwidths stated in the table are not the 3 dB
bandwidth as the video bandwidths are corrected for optimal flatness. Refer
to the E- Series E9320 power sensor Operating and Service Guide for
information on the sensor's peak flatness response. A filter
Off
mode is also
provided.
Accuracy
•
Instrumentation:Add the corresponding power sensor linearity percentage
(Refer to the Specifications section in the user's guide supplied with your
power sensor).
Table 1-1
Video Bandwidth vs. Peak Power Dynamic Range
Sensor
Video Bandwidth / Maximum Peak Power Dynamic Range
Off
High
Medium
Low
E9321A
300 kHz/
300 kHz/
100 kHz/
30 kHz/
E9325A
−
40 dBm to +20 dBm
−
42 dBm to +20 dBm
−
43 dBm to +20 dBm
−
45 dBm to +20 dBm
E9322A
1.5 MHz/
1.5 MHz/
300 kHz/
100 kHz/
E9326A
−
36 dBm to +20 dBm
−
37 dBm to +20 dBm
−
38 dBm to +20 dBm
−
39 dBm to +20 dBm
E9323A
5 MHz/
5 MHz/
1.5 MHz/
300 kHz/
E9327A
−
32 dBm to +20 dBm
−
32 dBm to +20 dBm
−
34 dBm to +20 dBm
−
36 dBm to +20 dBm
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com