Performance Tests
2
E4416A/E4417A Power Meters Service Guide
33
Procedure
1
Connect the equipment as shown in
Figure 2- 5
.
2
Unplug the range calibrator's power cord. Due to a ground loop design
problem with the 11683A Option H01 range calibrator, it does not operate
properly unless ground loops are eliminated. The range calibrator
continues to operate with no power applied, and disconnecting the power
cord eliminates the problem.
3
Figure 2- 5
shows the HI/LO outputs of the DC calibrator connected to the
DC reference input of the range calibrator. Ensure that the Guard and LO
output of the DC calibrator are floating, and are not connected to the earth
ground.
4
Switch the power meter on.
5
Preset the instrument:
Press
then .
6
Set the display units to watts:
Press
,
,
.
7
Set the display resolution:
Press
,
then press
until
is selected.
N O T E
The following procedure should be performed for the Instrument Accuracy Test.The
procedure details the key presses required on the Agilent E4416A. For the Agilent
E4417A the equivalent key presses should be performed on both channels.
WA RN I N G
There is a small risk of mild electrical shock from the chassis of the 11683A
Option H01 range calibrator unless the power meter is connected.
Preset
Local
Confirm
Meas
Display
Units
W
Meas
Display
Resolution
4