2
Performance Tests
34
E4416A/E4417A Power Meters Service Guide
8
Set the range calibrator as follows:
•
RANGE: 100 mW
•
POLARITY: NORMAL
•
FUNCTION: STANDBY
•
REF SWITCH: EXT
•
LINE: OFF
9
Set the DC calibrator as follows:
•
SET VALUE: 0 (ZERO)
•
FUNCTION: DC
•
OUTPUT: OFF
10
Set the filter length:
•
Press .
•
Highlight the
Filter
parameter using the up/down arrow keys.
•
Scroll through the
Filter
values by pressing
. Select a value of
MAN.
•
Highlight the filter length using the forward arrow key. Press
.
The Filter Length pop- up window is displayed. Set the filter length to
512.
•
Press .
11
Zero the power meter:
Press
,
. Wait approximately 10 seconds for the
wait symbol to disappear.
12
Set the DC calibrator output to 144.93 mV, and set the output switch to '+'.
N O T E
When switching the range calibrator to STANDBY, allow enough time for the range
calibrator to settle to its zero value before attempting to zero the power meter. This
settling would appear on the power meter display as downward drift. When the drift
has reached minimum, (typically less than 60 seconds), the range calibrator is settled.
Channel
Change
Change
Enter
Zero
Cal
Zero