Power Meter Operation
2
E4418B Power Meter User’s Guide
81
RS232/RS422
With RS232 selected as the serial interface, the power meter
communicates through the serial port with single ended drivers and
receivers. With RS422 selected, the serial port uses balanced drivers and
receivers.
The power meter acts as a DTE device.
To select RS232 or RS422 as the remote interface, press
,
,
,
, or
.
RS232/422 Connector
The serial port connector is a 9 pin male D- type wired as shown in
.
Figure 2-18
RS232/ 422 Pin Assignment
Setting the RS232/422 Parameters
This section explains how to set the baud rate, word size, parity, number
of stop bits, pacing, and echo for the serial interface.
Remote Interface
Select Interface
RS232
RS422
Pin
RS232
RS422
1
DCD
CTS-
2
Rx
Rx-
3
Tx
Tx+
4
DTR
Tx-
5
GND
GND
6
DSR
Rx+
7
RTS
RTS+
8
CTS
CTS+
9
RI
RTS-