Replaceable Parts
Assembly and Disassembly Guidelines
5-10
Agilent E4418B/E4419B Service Guide
Removing the A2 Processor Assembly
1.
Remove the A5 daughter and A6 measurement assemblies as described on
page 5-12.
2.
Move the A2 plastic support bracket to its forward position using the two
side levers, unclip the flexi-cable retaining bar on the front panel keypad and
front panel LCD cable connectors and disconnect the cable.
3.
Disconnect the following cables from the A2 processor assembly:
n
power reference semi-rigid
n
fan connector
n
power supply connector
4.
Turn the power meter upside down and remove the three screws as shown.
(When replacing these screws use a 6 lb/in T10 screw driver.)
5.
Push the A2 processor assembly towards the front panel to release it from
the 50-pin connector. Lift it upwards to remove.
Note
When re-assembling the processor board, ensure the A2 plastic support
bracket is returned to its locked position.
Note
After replacing a processor board, the Power Reference Frequency and
Level must be checked. For further information refer to “Performance
Tests”, on page 2-1.
Note
After replacing a processor board, the display brightness and contrast must
be adjusted. See “Adjustments”, on page 3-1.
Note
Firmware should be downloaded to the instrument after the processor board
is replaced. Refer to “Downloading Firmware”, on page 5-7.
screws
4402serv.book Page 10 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...