Theory of Operation
A2 Processor Assembly
Agilent E4418B/E4419B Service Guide
4-5
The LCD controller on the A3 front panel assembly is configured as a
memory mapped peripheral, and as such requires only to be fed with the
appropriate address, data and control lines from the microcontroller
circuits. The bias voltage for the LCD is produced by a DC to DC converter
that takes the +5 V (DIST) voltage and converts that to a n21 V.
The DC to DC converter is adjusted by a combination of the contrast
control signal from the microcontroller and the temperature sense voltage
that is generated on the A3 front panel assembly. The temperature sense
voltage helps to compensate for the normal variation of LCD contrast with
temperature. Current to the LCD’s LED backlight array is sourced from a
constant current circuit that compensates for variation in LED knee
voltage. A control input to this circuit from the front panel temperature
sense voltage allows for de-rating of the maximum LED current at high
operating temperatures. A control input from the microcontroller allows
the optimum backlight brightness to be factory set. The temperature
sense voltage, LED current and LCD bias voltage are all measured at self
test.
An analog multiplexer provides means by which several diagnostic points
can be switched to the A6 measurement assembly(s) for analogue to
digital conversion. The microcontroller uses this function to perform self
test.
A6 measurement assembly(s) to microcontroller communication is
conducted on a serial interface bus, the Internal Serial Bus (ISB). The ISB
is connected to the bi-directional serial port on the microcontroller via
buffers. The system clock and some control signals are connected to the
measurement modules. The measurement module interface is completed
with the distribution of switched and direct power (+12 V, -12 V and +5 V).
The TTL outputs are connected to the A6 processor assembly via in line
resistors located on the A4 interface assembly. A TTL logic level is output
from the processor assembly when the user defined windows based limits
are exceeded.
The TTL inputs are connected to the processor card via in line resistors
located on the interface card and can be used to initiate zero and
calibration cycles on the power meter.
4402serv.book Page 5 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...