Specifications
Power Meter Specifications
1-4
Agilent E4418B/E4419B Service Guide
Accuracy
Instrumentation
Absolute:
±
0.02 dB (Logarithmic) or
±
0.5% (Linear). (Refer to the power
sensor linearity specification in your power sensor manual to assess
overall system accuracy.)
Relative:
±
0.04 dB (Logarithmic) or
±
1.0% (Linear). (Refer to the power
sensor linearity specification in your power sensor manual to assess
overall system accuracy.)
Zero Set (digital settability of zero): Power sensor dependent (refer to
Table 1-1). For Agilent E-series power sensors, this specification applies
when zeroing is performed with the sensor input disconnected from the
POWER REF.
Table 1-1: Zero Set Specifications
Power Sensor
Zero Set
Agilent 8481A
±
50 nW
Agilent 8481B
±
50
µ
W
Agilent 8481D
±
20 pW
Agilent 8481H
±
5
µ
W
Agilent 8482A
±
50 nW
Agilent 8482B
±
50
µ
W
Agilent 8482H
±
5
µ
W
Agilent 8483A
±
50 nW
Agilent 8485A
±
50 nW
Agilent 8485D
±
20 pW
Agilent R8486A
±
50 nW
Agilent R8486D
±
30 pW
Agilent Q8486A
±
50 nW
Agilent Q8486D
±
30 pW
Agilent V8486A
±
200 nW
Agilent W8486A
±
200 nW
Agilent 8487A
±
50 nW
Agilent 8487D
±
20 pW
Agilent E4412A
±
50 pW
Agilent E4413A
±
50 pW
4402serv.book Page 4 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...