Performance Tests
Zero Test
Agilent E4418B/E4419B Service Guide
2-5
Zero Test
Specification
Description
After the power meter is initially calibrated and zeroed, the change in the
digital readout is monitored. This test also takes drift and noise into
account, since drift, noise and zero readings cannot be separated.
Equipment
Range Calibrator .................................. Agilent 11683A
Power Sensor Cable ............................. Agilent 11730A
Test Setup
Figure 2-1: Zero Test Setup
Electrical Characteristics
Performance Limits
Accuracy: Zero set (Digital settability of zero)
±
0.0764
µ
W
1
1. This performance limit is determined by the zero set specification of the power
sensor used in the measurement plus the measurement noise. The range
calibrator has a zero set specification of
±
0.05
µ
W. The calibrator measurement
noise specification is 110 nW at 16 averages. At 512 averages a noise multiplier of
0.24 is required, giving a measurement noise specification of 0.0264
µ
W
(0.24x512).
Agilent 11730A
POWER
METER
CHANNEL A
Power Meter
Range
Calibrator
4402serv.book Page 5 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...