Performance Tests
Instrument Accuracy Test
Agilent E4418B/E4419B Service Guide
2-7
Instrument Accuracy Test
Specification
Description
The power meter accuracy is verified for various power inputs. There are
two different methods described here that can be used to perform this test,
although one of these methods has a relatively high measurement
uncertainty values associated with it.
Test Procedure 1
Note
The measurement uncertainty of this test procedure is relatively
high at power levels of 30uW and below, as it uses a standard
11683A Range Calibrator. As such, this test procedure should be
used for indication only.
Equipment
Range Calibrator ................................... Agilent 11683A
Power Sensor Cable............................... Agilent 11730A
Electrical Characteristics
Performance Limits
Accuracy
±
0.5% or ±0.02 dB
1
1. This performance limit does not include the corresponding sensor power
linearity specification.
4402serv.book Page 7 Tuesday, October 14, 2003 3:18 PM