Performance Tests
Performance Test Record
2-30
Agilent E4418B/E4419B Service Guide
Performance Test Record
Model Agilent E4418B/E4419B Power Meter
Tested by ___________________________________________
Serial Number _______________________________________
Date _______________
Table 2-8: Performance Test Record
Test
Min
Result
Channel A
Actual
Result
Channel B
Actual
Result
Max
Result
Zero Test
−
76.40 nW
+
76.40 nW
Instrument Accuracy
3.1623
µ
W
10
µ
W
31.6228
µ
W
100
µ
W
316.227
µ
W
1 mW
3.1623 mW
10 mW
31.6228 mW
100 mW
3.0977
µ
W
9.904
µ
W
31.428
µ
W
99.49
µ
W
314.72
µ
W
0.9954 mW
3.1477 mW
9.954 mW
31.477 mW
99.54 mW
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
3.2269
µ
W
10.096
µ
W
31.819
µ
W
100.5
1 µ
W
317.74
µ
W
1.0047 mW
3.1769 mW
10.046 mW
31.769 mW
100.46 mW
Power Reference
P
meas
0.999
mW
___________
___________
1.001 mW
Interface Test
Pass/Fail
___________
___________
Pass/Fail
Functional Test
Low Range, 30 dBm pad
Low Range, 30 dBm pad
Low Range, no pad
Upper Range, no pad
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
___________
___________
___________
___________
___________
___________
___________
___________
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
4402serv.book Page 30 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...