Power Meter Operation
2
E4418B Power Meter User’s Guide
61
Making Relative Measurements
Relative mode allows you to compare any measurement result to a
reference value. The relative reading can be displayed in either dB or % by
pressing
and selecting either
or
. When the measurement
result is displayed in %, a prefix multiplier may be shown.
Relative mode is window dependent and when enabled, displays “Rel” in
the window to which it’s applied.
Procedure
To set a reference value on the currently selected window:
1
Press
.
2
Press
to use the current reading as the reference value. This allows
you to compare any measurement result in dB or percentage (%).
3
is automatically set to “On”, when
is pressed.
Successive measurements are now displayed relative to the reference
value. The relative mode can be disabled and re- enabled simply by
pressing
.
dB
%
Rel
Rel Off On
Rel
Rel Off On
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...