Chapter 2
19
Performance Test
Individual Test Description
3. Ch
apter
T
itle
4. Chapter
T
itle
5. Chapter T
itle
2. Pe
rfor
ma
nce T
e
st
Individual Test Description
1. Frequency Accuracy Test
Description
This test checks the frequency accuracy of the E4980A test signal. The frequency accuracy
is checked at 1MHz with a frequency counter. Since the E4980A employs a Direct Digital
Synthesizer for the signal source, the frequency accuracy test at this frequency point can
verify the accuracy for the entire frequency range.
2. DC Source Level Accuracy Test
Description
This test checks the DC source level accuracy of the E4980A DC source port output signal.
3. DC Bias Level Accuracy Test
Description
This test checks the DC bias level accuracy of the E4980A DC bias signal.
4. DC Bias V Level Monitor Accuracy Test
Description
This test checks the DC bias V level monitor accuracy of the E4980A. (Option : 001 only)
5. DC Bias I Level Monitor Accuracy Test
Description
This test checks the DC bias I level monitor accuracy of the E4980A. (Option : 001 only)
6. AC Signal Level Monitor Accuracy Test
Description
This test checks the AC signal level monitor accuracy of the E4980A.
7. AC Signal Level Accuracy Test
Description
This test checks the AC signal level accuracy of the E4980A output signal.
Summary of Contents for E4980A
Page 12: ...12 Contents ...
Page 60: ...60 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 86: ...86 Chapter5 Replaceable Parts Replaceable Parts List ...
Page 126: ...126 Chapter6 Replacement Procedure Display Interface Replacement ...
Page 136: ...136 AppendixB Firmware Update Update the E4980A firmware using USB cable ...
Page 140: ...140 AppendixC Power Requirement Power Requirements Figure C 3 Power Cable Supplied ...
Page 152: ...152 AppendixD Error Messages Error number 45 ...