Chapter 4
33
Troubleshooting
Troubleshooting Using Service Function
4. T
roub
lesho
o
ting
Figure 4-3
SELF TEST Page
Choosing a Test Item
Functional Description
E4980A can run the following self tests:
The tests of SYSTEM, USER DATA, and BATTERY are also performed by the power on
test.
㪼㪋㪐㪏㪇㪸㫌㪼㪈㪈㪋㪍
㪑㩷㩷㪝㫀㪼㫃㪻
Table 4-2
Test item
Description
SYSTEM
Checks the entire system, A1/A2/A3 boards, system
calibration data, and so on.
USER DATA
Checks the GPIB/LAN configurations, instrument
configuration information, instrument calibration data, and
scanner calibration data.
BATTERY
Checks the internal batteries.
KEY
Checks the font panel keys. (Visual checks only; no result
shown on screen.)
DISPLAY
Checks the font panel LEDs and LCD. (Visual checks only; no
result shown on screen.)
BIAS INTERFACE
Checks the bias interface. (Visual checks only; no result shown
on screen.)
Summary of Contents for E4980A
Page 12: ...12 Contents ...
Page 60: ...60 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 86: ...86 Chapter5 Replaceable Parts Replaceable Parts List ...
Page 126: ...126 Chapter6 Replacement Procedure Display Interface Replacement ...
Page 136: ...136 AppendixB Firmware Update Update the E4980A firmware using USB cable ...
Page 140: ...140 AppendixC Power Requirement Power Requirements Figure C 3 Power Cable Supplied ...
Page 152: ...152 AppendixD Error Messages Error number 45 ...