Appendix D
145
Error Messages
Error number: 22
D.
Er
ro
r M
e
ssa
g
e
s
error. When this error occurs, contact Agilent Technology’s Sales and Service Office or the
company from which you bought the device.
22
Fixture over temperature
The temperature inside the 42842A/B bias test fixture exceeds the limit. Turn off the DC
bias and cool down the equipment.
30
Function type not available
The Ls-Rdc, Lp-Rdc and Vdc-Idc are not available when the 42841A is connected.
G
-105
GET not allowed
GET is not allowed inside a program message.
H
70
Handler I/F not installed
The handler interface function is set to ON in a model that does not have a handler
interface board (option 201).
-114
Header Suffix out of Range
The header suffix is out of range.
I
73
I BIAS I/F not installed
The DC bias interface function is set to ON in a model that does not have the DC bias
interface board (option 002).
-224
Illegal parameter value
The parameter value is not properly set. This error occurs, for example, when the
:FUNC:IMP command is used to specify an Impedance-parameter that does not exist (as
CPRS).
51
Inconsistent limit setting
For the comparator function, user attempts to set the threshold in the sequential mode by
sending the COMP:SEQ:BIN command, although the tolerance mode is now selected.
Select the sequential mode using COMP:MODE.
-161
Invalid block data
Invalid block data was received (for example, END received before length satisfied).
-101
Invalid character
Invalid character was received.
-141
Invalid character data
Bad character data or unrecognized character data was received.
-121
Invalid character in number
Invalid character in numeric data.
Summary of Contents for E4980A
Page 12: ...12 Contents ...
Page 60: ...60 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 86: ...86 Chapter5 Replaceable Parts Replaceable Parts List ...
Page 126: ...126 Chapter6 Replacement Procedure Display Interface Replacement ...
Page 136: ...136 AppendixB Firmware Update Update the E4980A firmware using USB cable ...
Page 140: ...140 AppendixC Power Requirement Power Requirements Figure C 3 Power Cable Supplied ...
Page 152: ...152 AppendixD Error Messages Error number 45 ...