148
Appendix D
Error Messages
Error number: 16
R
16
Reference Measurement aborted
Reference data measurement is aborted.
S
71
Scanner I/F not installed
Operation that requires option 301 is performed in a model that does not have a scanner
interface board (option 301).
82
Store failed
User attempts to save data into number 10 to 19 when a USB memory device is not
inserted. Prepare the USB memory device or save data into number 0 to 9.
-150
String data error
String data is improper.
-158
String data not allowed
String data is not allowed.
-138
Suffix not allowed
A suffix is not allowed for this operation.
-102
Syntax error
Unrecognized command or data type was received.
T
-223
Too many digits
The received numeric value is too long (exceeds 255 digits).
-223
Too much data
The block-, expression-, or character-string-type program data that are received conform
with the standard but exceed the amount that can be processed under the condition of the
memory or conditions specific to memory-related devices. In this instrument, this error
occurs when the number of characters exceeds 254 in a character-string parameter.
-211
Trigger ignored
The instrument receives and detects a trigger command (“:TRIG”) or an external trigger
signal, but it is ignored due to the timing condition (the instrument is not in the wait-for-
trigger state, for example). Change the setup so that a trigger command or an external
trigger signal can be sent after the instrument has entered the wait-for- trigger state.
U
-113
Undefined header
A command not defined in this instrument, though not illegal in the syntactic structure, has
been received. For example, when the message “:OUTP:DC:ISOL:FREQ:AUTO ON” is
sent instead of the correct program message “:OUTP:DC:ISOL:LEV:AUTO ON” the
message sent is received as an undefined command by this instrument. See the command
Summary of Contents for E4980A
Page 12: ...12 Contents ...
Page 60: ...60 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 86: ...86 Chapter5 Replaceable Parts Replaceable Parts List ...
Page 126: ...126 Chapter6 Replacement Procedure Display Interface Replacement ...
Page 136: ...136 AppendixB Firmware Update Update the E4980A firmware using USB cable ...
Page 140: ...140 AppendixC Power Requirement Power Requirements Figure C 3 Power Cable Supplied ...
Page 152: ...152 AppendixD Error Messages Error number 45 ...