116
Chapter 5
Display Setting
Trace Comparison Using Memory Trace
Trace Comparison Using Memory Trace
Comparison with a stored reference trace
Procedure
Step 1.
Measure and display the reference trace.
Step 2.
Click inside the window of the reference trace (or click the measurement parameter name
area (e.g.
2:
θ
z [
°
])
in overlay display) (or press the
key) to make the trace active.
Step 3.
Right-click to open the shortcut menu and select
Display
(or press the
key).
Step 4.
Click the
Copy Data
→
Memory
button to store the reference trace’s data in the memory
trace.
NOTE
When the
Copy Data
→
Memory
button is clicked, usually only the active trace’s data are
stored in the memory trace. However, if the active trace is one of different overlaid traces
in the same window, the data of all traces in the window are stored in memory traces.
Step 5.
Click on the
Define Trace
box and select the method of displaying the trace.
where
is the measurement data and
is the data stored by the
Copy Data
→
Memory
button.
Subtraction of Offset Value
The result of subtracting an offset value from the measurement value can be displayed as a
data trace.
Procedure
Step 1.
Click inside the window of the trace from which the offset value will be subtracted (or
Define Trace
Box
Trace Display Method
Data
Data Trace:
Memory Trace: (No display)
Memory
Data Trace: (No display)
Memory Trace:
Data & Memory
Data Trace:
Memory Trace:
Data
−
Memory
Data Trace:
Memory Trace: (No display)
Delta %
Data Trace:
Memory Trace: (No display)
DATA
MEM
DATA
MEM
DATA MEM
–
DATA MEM
–
MEM
------------------------------------ 100
×
DATA
MEM
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...