Chapter 9
211
Setup and Use of Control/Management Functions
System Recovery
9. Setup
a
nd Use of
Con
trol/
Man
a
ge
me
nt F
unction
s
Procedure to execute the user recovery function
Returns the contents of the C drive to a user-specified state. To use this function, you must
create the user backup image in advance. For more information, see the description
“Procedure to create the user backup image” on page 208.
NOTE
This function is available when the volume label on the hard disk is IG201 or higher.
NOTE
You need the keyboard for this operation.
Step 1.
Shut down the E4991A.
Step 2.
Connect the keyboard to the E4991A.
Step 3.
Insert the disk for the system recovery into the floppy disk drive of the E4991A.
Step 4.
Press the standby switch of the E4991A to turn it on.
Step 5.
When the screen as shown in the figure below appears, press and hold
of the
keyboard until this screen disappears.
NOTE
After several seconds, the next screen appears automatically even if you do not press any
key, so do not miss it.
If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...