Chapter 11
239
Specifications and Supplemental Information
Source Characteristics
11
. S
pecifications an
d
Su
pple
m
en
ta
l Infor
m
ation
Source Characteristics
Frequency
Oscillator Level
Range
1 MHz to 3 GHz
Resolution
1 mHz
Accuracy
without Option 1D5
±
10 ppm (23
±
5
°
C)
±
20 ppm (5
°
C to 40
°
C)
with Option 1D5
±
1 ppm (5
°
C to 40
°
C)
Stability
with Option 1D5
±
0.5 ppm/year (5
°
C to 40
°
C) (Typical)
Range
Power (when 50
Ω
load is
connected to test port)
−
40 dBm to 1 dBm (Frequency
≤
1 GHz)
−
40 dBm to 0 dBm (Frequency > 1GHz
*1
)
*1. It is possible to set an oscillator level of more than 0 dBm (447 mV, 8.94 mA) at
frequency > 1 GHz. However, the characteristics at this setting are not guaranteed.
Current (when short is
connected to test port)
0.0894 mArms to 10 mArms (Frequency
≤
1 GHz)
0.0894 mArms to 8.94 mArms (Frequency > 1 GHz
Voltage (when open is
connected to test port)
4.47 mVrms to 502 mVrms (Frequency
≤
1 GHz)
4.47 mVrms to 447 mVrms (Frequency > 1 GHz
)
Resolution
0.1 dB
*2
*2. When the unit is set at mV or mA, the entered value is rounded to 0.1 dB resolution.
Power Accuracy
Test Head
*3
*3. When 50
Ω
load is connected to test port of test head.
Frequency
≤
1 GHz
±
2 dB (23
±
5
°
C)
±
4 dB (5
°
C to 40
°
C)
Frequency > 1 GHz
±
3 dB (23
±
5
°
C)
±
5 dB (5
°
C to 40
°
C)
Option 010 Probe Station Connection Kit
Frequency
≤
1 GHz
±
5.5 dB (5
°
C to 40
°
C)
Frequency > 1 GHz
±
7.6 dB (5
°
C to 40
°
C)
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...