312
Appendix A
Manual Changes
Manual Changes
a.
Connect the E4991A to a LAN with a LAN cable after the power to the E4991A has
been turned off.
b.
Press the Standby switch once to reset the switch and then press the switch once more
to turn the power on.
Change 4
When the serial number prefix is JP2KH and MY432, change the description of “System
Recovery” on page 204 to the following one.
System Recovery
Performing system recovery will allow you to reset the Windows operating system and the
firmware of the E4991A to the state when you purchased
*1
.
If, for some reason, the Windows operating system or firmware failed and cannot be started
normally or operation after startup is unstable, execute system recovery.
Notes on executing system recovery
Performing system recovery causes the following:
o The following settings of the E4991A are initialized.
•
Network settings
•
GPIB settings
•
Printer settings
o If the firmware has been updated after purchasing the E4991A, the firmware when you
purchased the product
o If you installed any driver software for the supported printer, it will be deleted.
Files you created using the save function are not affected, but we recommend backing them
up before executing system recovery for precautionary purposes.
Procedure to execute system recovery
NOTE
You need the keyboard for this operation.
Step 1.
Shut down the E4991A.
Step 2.
Connect the keyboard to the E4991A.
Step 3.
Press the standby switch of the E4991A to turn it on.
Step 4.
When the message in the figure below appears on the screen following the Agilent's logo
screen, immediately press
on the keyboard.
*1.If the hard disk failed and has been replaced after purchase, the state when the
replacement was performed is recovered.
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...