324
Appendix B
Probe Station Connection Kit (Option 010)
Mounting Test Head and Connecting Cables (using recommended probe
station)
Mounting Test Head and Connecting Cables
(using recommended probe station)
To mount the test head, you need the mounting plate and the semi-rigid cable provided by
Cascade Microtech in addition to Option 010. The mounting plate is used to connect the
test head to the probe arm. The semi-rigid cable is used to connect the test head to the
probe head. Their are two types of semi-rigid cables: one for the Summit 9000 series and
another for the Summit 11000/12000 series. Select the appropriate type for your probe
station. To order these parts, please inquire to Cascade Microtech.
Step 1.
Fix the test head to the mounting plate supplied by Cascade Microtech.
Figure B-2
Fixing Test Head to Mounting Plate
Step 2.
Connect the 3.5-mm to 7-mm adapter to the test head’s 7-mm connector.
Summary of Contents for E4991A