Temperature compensation
Executing the temperature compensation feature will reduce an error due to temperature
change. The temperature compensation feature uses reference data to compensate an error
that may be related to the measuring cable exposed to temperature change. Prior to
compensation, the reference data is prepared by obtaining variation of measurement values
of the open/short standards (heat-resistant) relative to normal temperature. The data will be
obtained for all temperatures to be measured.
The temperature compensation is executed using a built-in VBA program called
TemperatureCompensation.
NOTE
You cannot execute the temperature compensation using the front panel.
You can perform temperature compensation more easily using the attached sample
program (tctest) that performs temperature compensation. In particular, when you use
ESPEC SU-261 as a temperature chamber, you can use the attached sample program
without any modification. For more information, refer to “Measuring temperature
characteristic using sample program” on page 351.
Execution procedure of temperature compensation
Follow the temperature compensation flow shown below.
Step 1.
Acquire temperature compensation data for the temperature you want to measure and save
it in a file. For more information, refer to “Acquiring temperature compensation data” on
page 344.
Step 2.
Execute measurement at a desired temperature.
Step 3.
Stop the sweep.
Step 4.
Execute the temperature compensation with the measurement result using the program for
compensation. Refer to “Information to create program to execute temperature
compensation” on page 341 to create a program to execute temperature compensation
using Temperature Compensation.
Information to create program to execute temperature compensation
Preparation for using TemperatureCompensation
TemperatureCompensation is included in the library called Agilent E4991A-007
Compensation Library (DLL). In order to use the Agilent E4991A-007 Compensation
Library, you need to enable the reference to it.
Enable the reference to Agilent E4991A-007 Compensation Library
Step 1.
Run the VBA editor.
Step 2.
On the
Tools
menu of the VBA editor, click
References...
to display the
References
window.
Step 3.
Check the box associated with Agilent E4991A-007 Compensation Library.
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...