Appendix C
359
Temperature Characteristic Test Kit (Option 007)
Measuring temperature characteristic using sample program
C. T
e
m
per
atur
e
Char
acter
istic
Te
st Kit (
O
ption
007
)
Figure C-19
Temperature Profile File (example)
NOTE
Since the recommended temperature chamber does not provide humidity control, humidity
is not specified in Figure C-21.
NOTE
No space is required between a value and a comma(,).
When you do not specify humidity, place no space between commas.
Always enter temperatures and waiting times.
NOTE
You can modify the program to change the limit values. For information on how to change
them, refer to “Changing limits when setting temperature change pattern” on page 371.
Step 2.
Setting a state file
Click the
State Files
button on the Main Menu (Figure C-13) to display Figure C-20.
Click the
Browse
button, choose a desired state file you saved in advance, and click the
OK
button.
You can specify up to 3 files each for A to C. Specify at least one file.
Table C-2
Unit, resolution, and limit values of temperature profile data
Unit
Resolution
Maximum
value
Minimum
value
Temperature
°
C
0.1
150
-55
Humidity
%
0.1
99
0
Waiting Time
Minute
1
999
1
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...