454
Appendix H
Comparison Information of 4291B and E4991A
Function Comparison List
Function Comparison List
Table H-1 compares the functions of the 4291B with those of the E4991A.
Table H-1
Function Comparison List
4291B
E4991A
Measure-
ment per-
formance
Frequency Range
1 MHz to 1.8 GHz
1 MHz to 3 GHz
Resolution
1 mHz
1 mHz
OSC
level
Voltage ranges
(in open condition)
0.2 mVrms to 1 Vrms (@
≤
1 GHz)
0.2 mVrms to 0.5 Vrms (@>1 GHz)
4.47 mVrms to 502 mVrms (@
≤
1 GHz)
4.47 mVrms to 447 mVrms (@>1 GHz)
Electric current ranges
(in short condition)
4
μ
Arms to 20 mArms (@
≤
1 GHz)
4
μ
Arms to 10 mArms (@>1 GHz)
89.4
μ
Arms to 10 mArms (@
≤
1 GHz)
89.4
μ
Arms to 8.4 mArms (@>1 GHz)
Power ranges (in 50
Ω
end
terminal condition)
-67 dBm to 7 dBm (@
≤
1 GHz)
-67 dBm to 1 dBm (@>1 GHz)
-40 dBm to 1 dBm (@
≤
1 GHz)
-40 dBm to 0 dBm (@>1 GHz)
DC
Bias
Voltage range
0 V to
±
40 V
0 V to
±
40 V
Electric current ranges
-100 mA to -20
μ
A
20
μ
A to 100 mA
-50 mA to -100
μ
A
100
μ
A to 50 mA
Basic accuracy
0.8%
0.8%
Impedance measurement range
100 m
Ω
to 40 k
Ω
(@ 1 MHz, <10%)
130 m
Ω
to 20 k
Ω
(@ 1 MHz, < 10%)
Test
station
(head)
Cables
1.8 m
0 m
(Direct connection)
Terminals
7 mm
7 mm
Heads
High impedance type
Low impedance type
Only one type
Number of channels
2
No channel concept
Traces
Data trace
Memory trace
(multiple)
User trace
Data trace
Memory trace
Measure-
ment
parameter
Number of parameters that can be
displayed on screen
Maximum
2
Maximum
5
(scalar: 3, complex:2)
Selectable parameters
Impedance measurement: |Z|,
θ
z, R, X,
|Y|,
θ
y, G, B, |
Γ
|,
θ
Γ
,
Γ
x,
Γ
y, Cp, Cs,
Lp, Ls, Rp, Rs, D, Q
Dielectric/Magnetic material measurement
(option):
ε
r`,
ε
r", tan
δ
(
ε
), |
ε
r|,
μ
r`,
μ
r",
tan
δ
(
μ
), |
μ
r|
Impedance measurement: |Z|,
θ
z, R, X,
|Y|,
θ
y, G, B, |
Γ
|,
θ
Γ
,
Γ
x,
Γ
y, Cp, Cs,
Lp, Ls, Rp, Rs, D, Q
Dielectric/Magnetic material measurement
(option):
ε
r`,
ε
r", tan
δ
(
ε
), |
ε
r|,
μ
r`,
μ
r",
tan
δ
(
μ
), |
μ
r|
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...