50
Chapter 3
Setting Measurement Conditions
Measurement Waiting Time/Sweep Time Setting
Measurement Waiting Time/Sweep Time Setting
A time delay can be set for the period after the sweep is triggered but before the actual
sweep starts (sweep delay). Furthermore, a time delay can also be set for the period after a
signal is applied to a DUT but before actual measurement starts (point delay). This
function is useful, for example, when a certain period is required before the characteristics
of the DUT can become stable after a signal is applied. Another application of this function
is to observe the changes in a DUT’s impedance over time for a long span.
Figure 3-3 shows the definitions of sweep time, sweep delay time, point delay time, and
segment delay time.
Figure 3-3
Definition of measurement waiting time
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...