56
Chapter 3
Setting Measurement Conditions
Setting Sweep Area with Marker
Procedure
Step 1.
Place the active marker on the new sweep start, sweep stop, or sweep center. To set a sweep
span, place Marker R and Marker 1 (or 2-8) on the trace points corresponding to the new
sweep span and set the marker delta mode to delta or fixed delta. At this time, set Marker 1
(or 2-8) to active status (marker display:
∇
). For details on how to use the marker, see
Chapter 6, “Analysis of Measurement Results,” on page 127.
Step 2.
Right-click to open the shortcut menu and click
Marker
(or press
).
Step 3.
Click the
Marker To Menu
button.
Step 4.
Set the position (distance) of markers on a new sweep area by clicking any of the following
buttons for assigning marker values.
NOTE
After the sweep area is changed, no trace is displayed in the new sweep area until a new
measurement is taken.
Button Assigning Marker Value
Sweep Area Display Mode
Start
Assign the stimulus value of the active marker as
the sweep start value. As a result, the position of
the marker establishes a new sweep start point.
Stop
Assign the stimulus value of the active marker as
the sweep stop value. As a result, the position of
the marker establishes a new sweep stop point.
Center
Assign the stimulus value of the active marker as
the sweep center value. As a result, the position
of the marker establishes a new sweep center
point.
Delta To Span
For Marker R and the active marker (one of
Markers 1 to 8), the smaller marker stimulus
value is assigned as the sweep start value and the
larger one is assigned as the sweep stop value.
As a result, the sweep range between Marker R
and the active marker establishes a new sweep
span.
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...