68
Chapter 3
Setting Measurement Conditions
Sweeping a Plural Number of Sweep Areas under Different Conditions at
One Time (Segment Sweep)
To change the numeric value in a spin box, right-click in the box to open and use the
numeric entry dialog box, or use the
ENTRY/NAVIGATION
block keys on the front
panel.
To select an item in a box, click to open the box and then click to select the item.
The maximum number of segments in a table is 16. The maximum number of
measurement points in the entire table is 801.
e.
Repeat step c and step d to complete the required segments.
To modify one of the created segments, enter the segment number in the
Segment No.
spin box and edit the segment in accordance with step d.
NOTE
You can directly call up the setup box for changing an item’s value by clicking the current
value in the segment table on the screen (Figure 3-7).
More
button
This opens the toolbar showing the buttons
described below.
<<
button
This returns to the toolbar showing the buttons
described above.
Osc Level
spin box
This sets source level of segment.
Osc Unit
list box
This selects source level unit in segment sweep
from among
Power
,
Voltage
and
Current
.
*1
Bias Level
spin box
This sets DC bias level of segment.
Bias limit
spin box
This sets DC bias limit of segment.
Bias Source
list box
This selects either
Voltage
or
Current
as DC
bias mode in segment sweep.
*1.This is a setting common to all segments (cannot be set individually for each segment).
Segment Sweep Setting Button/Box
Setting Function
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...