Chapter 6
137
Analysis of Measurement Results
Selecting Marker Value Display for Plural Parameters
6.
Ana
lysis of
M
easur
em
ent
Result
s
Selecting Marker Value Display for Plural Parameters
When using a marker for plural traces, the marker value is displayed in the form of two
scalar parameters converted from two plural parameters (main and sub-parameters). This
conversion method enables selection from six kinds of marker values (Table 6-1),
irrespective of the coordination format set (polar coordination, plural planes, Smith chart,
Admittance chart).
Table 6-1
Marker Value Display Mode and Marker Value
Displayed
Marker Value Display
Mode (selection in
Smith/Polar
box)
Marker Value
Main Parameter
Sub-parameter
Real Imag
Actual number component
Imaginary number
component
LinMag Phase
Amplitude (linear)
Phase
LogMag Phase
Amplitude (log, dB unit)
Phase
R+jX
Resistance
(actual number component
of plural impedances
*1
)
*1. In case of measuring plural impedances (Z), the same value is displayed as when
Real
Imag
is selected.
Reactance
(Imaginary number
component of plural
impedances
G+jB
Conductance
(actual number component
of plural admittances
*2
)
*2. In case of measuring plural admittances (Y), the same value is displayed as when
Real
Imag
is selected.
Susceptance
(Imaginary number
component of plural
admittances
)
Swr Phase
Standing wave ratio
Phase of reflection
coefficient
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...