Chapter 6
139
Analysis of Measurement Results
Selecting Marker Analysis Target Trace (Data/Memory)
6.
Ana
lysis of
M
easur
em
ent
Result
s
Selecting Marker Analysis Target Trace (Data/Memory)
When the memory trace is used according to the procedures of “Trace Comparison Using
Memory Trace” on page 116, it is necessary to select the memory trace as the target of
analysis by the marker (data trace is the preset setting).
Procedure
Step 1.
Click inside the window of the trace to select it as the target trace for marker analysis (or
click the measurement parameter name area, e.g.
2:
θ
z [
°
]
in the case of overlay display) (or
press
) and make the trace active.
Step 2.
Right-click to open the shortcut menu and click
Marker
(or press
).
Step 3.
Click the
Marker On
button and select the marker analysis target trace.
Marker On
Button Display
Marker Analysis Target Trace
Marker On: [Data]
Data trace
Marker On: [Memory]
Memory trace