238
Chapter 11
Specifications and Supplemental Information
Definitions
Definitions
All specifications apply over a 5
°
C to 40
°
C range (unless otherwise stated) and 30 minutes
after the instrument has been turned on.
Supplemental information is intended to provide information that is helpful for using the
instrument but that is not guaranteed by the product warranty. This information is denoted
as either typical or nominal.
Measurement Parameters and Range
Measurement Parameters
Measurement Range
Specification (spec.):
Warranted performance. Specifications include guardbands to
account for the expected statistical performance distribution,
measurement uncertainties, and changes in performance due to
environmental conditions.
Typical (typ.):
Expected performance of an average unit that does not include
guardbands. It is not guaranteed by the product warranty.
Nominal (nom.):
A general, descriptive term that does not imply a level of
performance. It is not guaranteed by the product warranty.
Impedance parameters
|Z|, |Y|, L
s
, L
p
, C
s
, C
p
, R
s
(R), R
p
, X, G, B, D, Q,
θ
z
,
θ
y
, |
Γ
|,
Γ
x
,
Γ
y
,
θγ
Material parameters (option 002)
(see “Option 002 Material Measurement (typical)” on
page 263)
Permittivity parameters
,
,
,
Permeability parameters
,
,
,
Measurement range (|Z|)
130 m
Ω
to 20 k
Ω
(Frequency = 1 MHz,
Point averaging factor
≥
8,
Oscillator level =
−
3 dBm; =
−
13 dBm; or =
−
23 dBm,
Measurement Accuracy
≤
±
10%,
Calibration is performed at 23
°
C
±
5
°
C,
Measurement is performed at calibration temperature
±
5
°
C)
ε
r
ε
r
′ ε
r
″
δ
tan
μ
r
μ
r
′ μ
r
″
δ
tan
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...