276
Chapter 11
Specifications and Supplemental Information
Option 007 Temperature Characteristic Test Kit
Option 007 Temperature Characteristic Test Kit
This section contains specifications and supplemental information for the E4991A Option
007. Except for the contents in this section, the E4991A standard specifications and
supplemental information are applied.
Operation Temperature
Source Characteristics
Frequency
Oscillator Level
Measurement Accuracy (at 23
°
C
±
5
°
C)
Conditions
The measurement accuracy is specified when the following conditions are met:
NOTE
The heat-resistant cable must be kept at the same position throughout calibration and
measurement.
Range
-55
°
C to +150
°
C (at the test port of the heat-resistant cable)
Range
1 MHz to 3 GHz
Source power accuracy at the test port of the heat-resistant cable:
Frequency
≤
1 GHz
+2 dB/-4 dB (23
°
C
±
5
°
C)
+4 dB/-6 dB (5
°
C to 40
°
C)
Frequency > 1 GHz
+3 dB/-6 dB (23
°
C
±
5
°
C)
+5 dB/-8 dB (5
°
C to 40
°
C)
Calibration
Open, Short and Load calibration is completed at the test port
(7-mm connector) of the heat-resistant cable
Calibration temperature
Calibration is performed at an environmental temperature
within the range of 23
°
C
±
5
°
C. Measurement error doubles
when calibration temperature is below 18
°
C or above 28
°
C.
Measurement temperature range
Within
±
5
°
C of calibration temperature
Measurement plane
Same as calibration plane
Oscillator level
Same as the level set at calibration
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...