Temperature Characteristic Test Kit (Option 007)
Measuring temperature characteristic using sample program
C. T
e
m
per
atur
e
Char
acter
istic
Te
st Kit (
O
ption
007
)
4. Acquiring temperature compensation data
When you perform temperature compensation, you need to acquire temperature
compensation data according to the following procedure before measuring the DUT.
NOTE
If the DC bias feature (option) is provided, turn it off while acquiring temperature
compensation data.
Step 1.
Specify measurement temperature settings using the following temperature profile file.
Figure C-21 shows the temperature profile file provided at the factory. You need to acquire
temperature compensation data for all measurement temperatures, making any changes as
necessary.
D:\TCTEST\CompTemp.Tpr
Edit a file on your external PC. File transfer between the external PC and the E4991A is
performed using the FTP server function of the E4991A via LAN. For more information on
the FTP server ffunction, refer to Chapter 10, “Using LAN.”.
NOTE
The only valid temperature profile file name when acquiring temperature compensation is
D:\TCTEST\CompTemp.Tpr.
In the temperature profile file, each temperature (and humidity) at which you want to
acquire compensation data and waiting time after the specified temperature (humidity)
reached as specified are written in the order of:
{temperature},{humidity},{waiting time}
separated with a comma (,). Each temperature point is separated with a line feed.
Figure C-21
CompTemp.Tpr temperature profile file (factory-set)
NOTE
Since the recommended temperature chamber does not provide humidity control, humidity
is not specified in Figure C-21.
NOTE
No space is required between a value and a comma(,).
When you do not specify humidity, place no space between commas.
Always enter temperatures and waiting times.
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...