Stimulus - Cal/Compen -
Cal Menu[ ]
-Recover Cal/Comp
State
Recovers the calibration/fixture compensation
data and restores the instrument setups. For more
on the instrument setups, refer to Appendix G on
page 439.
None
Menu Bar
(Key)
Setup Toolbar
Function
GPIB Command
Menu Bar
(Key)
Setup Toolbar
Function
GPIB Command
Stimulus
-
Cal/Compen.
(
)
-Cal Menu[ ]
Calibration:
-Meas Open
Measures the calibration data in the OPEN
standard of the calibration kit. When the
measurement is completed, there is a check mark
(
√
) to the left of the button.
SENS:CORR1:COLL
-Meas Short
Measures the calibration data in the SHORT
standard of the calibration kit. When the
measurement is over, there is a check mark (
√
) to
the left of the button.
SENS:CORR1:COLL
-Meas Load
Measures the calibration data in the LOAD
standard of the calibration kit. When the
measurement is over, there is a check mark (
√
) to
the left of the button.
SENS:CORR1:COLL
-Meas Low Loss C
Measures the low loss capacitor calibration data
(attachment only) in the OPEN standard of the
calibration kit. When the measurement is over,
there is a check mark (
√
) to the left of the button.
Low loss capacitor calibration is only necessary
when High Q (low loss factor) measurement is
done at high precision.
SENS:CORR1:COLL
-Done
-(Abort Cal Meas)
Done: Available when the measurement of the
OPEN, SHORT, LOAD (low loss capacitor)
calibration data is completed. The calibration
coefficient is calculated based on the acquired 3
(4) calibration data, which are then stored into
memory to enable calibration.
(Abort Cal Meas): Available only in OPEN,
SHORT, LOAD (low loss capacitor) calibration
measurement. Aborts calibration measurement.
SENS:CORR1:COLL:SAVE
(None)
-Cal Reset
Disables all calibration data and the calibration
coefficient acquired. When the button is clicked,
the check mark (
√
) to the left of the button
disappears.
SENS:CORR1
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...