When you take a material measurement (permittivity or permeability measurement) with
the E4991A and “Option 002 material measurement,” the following setups must first be
completed.
Procedure
Step 1. Select the type of material measurement
a.
Right-click to open the shortcut menu and click
Utility
(or press
).
b.
Click the
Material Option Menu
button.
c.
Click to open the
Material Type
box and click on the material measurement type.
Step 2. Enter material dimensions
Using the numeric entry dialog box that appears by right-clicking inside the setting box
described below (or using the
ENTRY/NAVIGATION
block key of the front panel), enter
the necessary material dimensions:
Material Type
Box
Material Measurement Type
Impedance
Impedance measurement (measurement of
general impedance parameters except for
permittivity and permeability).
Permittivity
Permittivity measurement
Permeability
Permeability measurement
Material Dimension Setting Box
Value to Enter
Thickness
Thickness of dielectric material t [m] (see Figure
3-1)
Height
Height of magnetic material h [m] (see Figure
3-2)
Inner Diameter
Inner diameter of magnetic material b [m] (see
Figure 3-2)
Outer Diameter
Outer diameter of magnetic material c [m] (see
Figure 3-2)
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...