Chapter 3
61
Setting Measurement Conditions
Setting the Trigger Source To Start Sweep
3
. Set
ting
M
e
as
ur
em
e
nt
Co
ndition
s
Setting the Trigger Source To Start Sweep
Set up the signal source to execute sweep start (i.e., the trigger source) by following the
procedure below.
Procedure
Step 1.
Right-click to open the shortcut menu and click
Trigger Setup
(or press
).
Step 2.
Click to open the
Trigger Source Box
and select the desired trigger source.
NOTE
For more details on the characteristics of the signal for an external trigger, see “Chapter 11,
“Specifications and Supplemental Information,” on page 237.”
For more details on using GPIB, see
Programming Manual
.
Trigger Source
Box
Trigger Source
Internal
Internal continuous trigger generated
automatically by the E4991A.
Manual
Manual trigger generated by clicking the
Manual
Trigger
button on the
Trigger Setup
toolbar or
by pressing
*1
.
*1.The manual trigger can also be executed by clicking
Trigger-Trigger
from the menu bar
or by clicking
Trigger
from the shortcut menu displayed by pressing the right button.
External
External trigger implemented upon detecting
TTL signal coming into external trigger input
connector of rear panel.
GPIB Bus
GPIB trigger implemented by trigger command
to start measurement sent via GPIB.
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...