Chapter 4
87
Calibration and Compensation
Electrical Length Compensation
4. Cal
ibra
tion
an
d
Comp
ensatio
n
Electrical Length Compensation
The electrical lengths of the test fixtures shown in Table 4-6 have been registered in the
E4991A in advance. To use these fixtures, you may set the needed electrical length by
simply selecting the model number of the fixture to be used. If you use fixtures that are not
registered, however, you must enter the electrical length as a numerical value.
NOTE
The 16453A is automatically selected when the permittivity measurement (
Permittivity
) is
selected as material type (
Material Type
).
The 16454A (fixture size: S) or 16454A (fixture size: L) can be selected when the
permeability measurement (
Permeability
) is selected as material type (
Material Type
).
For selecting the material type, refer to “Setting Material Measurement Parameter (Option
002 only)” on page 43.
Table 4-6
Test fixtures with registered electrical lengths
Model Number
Electrical Length
16191A
14 mm
16192A
11 mm
16193A
14 mm
16194A
50 mm
16196A
26.2 mm
16196B
26.9 mm
16196C
27.1 mm
16197A
14 mm
16453A
0 mm
16454A (Fixture size: S)
0 mm
16454A (Fixture size: L)
0 mm
Summary of Contents for E4991A
Page 6: ......
Page 24: ...18 Contents ...
Page 80: ...74 Chapter3 Setting Measurement Conditions Averaging Measurement Results ...
Page 220: ...214 Chapter9 Setup and Use of Control Management Functions System Recovery ...
Page 338: ...332 AppendixB Probe Station Connection Kit Option 010 OPEN SHORT LOAD Calibration ...
Page 428: ...422 AppendixD Menu References Menu References ...
Page 482: ...476 AppendixI Messages Wait measuring comp standard ...