Chapter 2
Performance Tests
Individual Test Description
2. Frequency Measurement Test
Description
This test checks the frequency measurement accuracy of the E5052A. The frequency
measurement is checked for each frequency band and each frequency resolution of the
E5052A
Specification
@23
±
5
°
C
Test equipment
Procedure for Standard
Step 1.
Connect the power sensor to the power meter. Perform the zero adjustment and the
calibration for the power meter.
Step 2.
Connect the test equipment as shown in Figure 2-2
Frequency accuracy
±
(reso time base accuracy) @ 10 MHz to 7 GHz
Frequency Resolution: 10 Hz, 1 kHz, and 64 kHz
Internal time base stability:
±
5 ppm
Signal Generator
Agilent E8257D
RF Amplifier (AMP 1)
R&K A3000-2H-R
RF Amplifier (AMP 2)
Agilent 83020A
Power Meter
Agilent E4419B
Power Sensor
Agilent 8481A
Step Attenuator
Agilent 8496H
Power Splitter
Agilent 11667A
Blocking Capacitor
Agilent 11742A
N(f)-SMA(m) adapter, 2 ea.
Agilent p/n 1250-1562
3 dB Fixed Attenuator, 2 ea.
Agilent 8491A Opt.003
N(m)-N(m) cable 61cm, 3 ea.
Agilent N6314A (p/n 8120-8862)
N(m)-N(m) Adapter
Agilent p/n 1250-1475
BNC cable, 122cm
Agilent p/n 8120-1840
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...