42
Chapter 2
Performance Tests
Individual Test Description
Step 17.
Turn off the power of AMP2.
Procedure for Option 011
Step 1.
Connect the power sensor to the power meter. Perform the zero adjustment and the
calibration for the power meter.
Step 2.
Connect the test equipment as shown in Figure 2-8
Figure 2-8
Power Measurement Accuracy Test setup 1 for Option 011
Step 3.
Press
[Preset]
-
Factory
to initialize the E5052A, then set the E5052A as follows. In case
of installing the firmware revision 1.00/1.10, press
[Preset]
-
OK
.
Step 4.
Preset the Signal Generator and Power Meter.
Step 5.
Turn on the power of AMP1.
Step 6.
Set the E5052A as follows.
E5052A settings
Key strokes
Measurement View : FP-Power
[Meas/View]
-
Freq & Power
Trigger Mode : Hold
[Trigger]
-
Trigger to Freq & Power
-
Hold
E5052A settings
Key strokes
Frequency Band : 10 M - 1.5
GHz
[Setup]
-
Frequency Band
-
10M-1.5GHz
Attenuator : 0 dB
[Attn]
-
Input Attenuator
-
[0]
-
[x1]
Trigger Mode : Continuous
[Trigger]
-
Continuous
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...