Chapter 2
59
Performance Tests
Individual Test Description
2. Pe
rfor
ma
nce T
e
st
7. DC Power Current Measurement Test
Description
This test checks the accuracy of the DC Power current measurement function. This test
uses a 141
Ω
resistor to define the DC Power current.
Specification
@23
±
5
°
C
Test Equipment
Procedure for Standard
Step 1.
Connect the test equipment as shown in Figure 2-13
Figure 2-13
DC Power Current Measurement Test setup for Standard
Step 2.
Press
[Preset]
-
Factory
to initialize the E5052A, then set the E5052A as follows. In case
of installing the firmware revision 1.00/1.10, press
[Preset]
-
OK
.
Accuracy
±
(0.2 % of r 160
μ
A) @ 0 mA to 80 mA
Multimeter
Agilent 3458A
Resistor 141
Ω
, 1 W (Agilent p/n 0698-3615 (47
Ω
), 3ea)
BNC cable, 61 cm, 2 ea.
Agilent p/n 8120-1839
Dual Banana-BNC(f) adapter
Agilent p/n 1251-2277
E5052A settings
Key strokes
Measurement View :
FP-Current
[Meas/View]
-
Freq & Power
Mode : Tester
[Trigger]
-
Mode
-
Tester
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...