Chapter 2
61
Performance Tests
Individual Test Description
2. Pe
rfor
ma
nce T
e
st
repeat Step 4 through 10.
Procedure for Option 011
Step 1.
Connect the test equipment as shown in Figure 2-14
Figure 2-14
DC Power Current Measurement Test setup for Option 011
Step 2.
Press
[Preset]
-
Factory
to initialize the E5052A, then set the E5052A as follows. In case
of installing the firmware revision 1.00/1.10, press
[Preset]
-
OK
.
Table 2-16
DC Power Current Test Point for Standard
Multimeter E5052A
Current Through
Resistor
Setting
Range
Setting
Voltage
2 mA
10 mA
0.3 V
5 mA
10 mA
0.7 V
10 mA
10 mA
1.41 V
20 mA
100 mA
3.0 V
50 mA
100 mA
7.0 V
80 mA
100 mA
11.3 V
E5052A settings
Key strokes
Measurement View : FP
[Meas/View]
-
Freq & Power
Trigger Mode : Hold
[Trigger]
-
Trigger to Freq & Power
-
Hold
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...