70
Chapter 2
Performance Tests
E5052A Performance Test Record
SSB Phase Noise Sensitivity Test (for Standard)
CW Frequency: 70 MHz, Correlation = 1, Start frequency = 1 Hz
SSB Phase Noise Sensitivity Test (for Option 011)
CW Frequency: 70 MHz, Start frequency = 10 Hz
SA Relative Level Accuracy Test
CW Frequency: 100 MHz
2.0125 G
±
0.5
±
0.13
3.0125 G
±
1.0
±
0.13
4.0125 G
±
1.0
±
0.16
5.0125 G
±
1.0
±
0.16
6.0125 G
±
1.0
±
0.17
7 G
±
1.0
±
0.19
Frequency [Hz]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
Offset Frequency
[Hz]
Test limit [dB/Hz]
Test result
[dB/Hz]
1 M
< -168.5
3 M
< -169.0
7 M
< -169.7
Offset Frequency
[Hz]
Test limit [dB/Hz]
Test result
[dB/Hz]
1 M
< -160.0
3 M
< -160.0
7 M
< -160.0
Offset Frequency
[Hz]
Test limit [dB]
Test result [dB]
Measurement
uncertainty [dB]
-7.5 M
±
1.5
±
0.17
-6.5 M
±
1.5
±
0.17
-5.5 M
±
1.5
±
0.16
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...