Chapter 4
99
Troubleshooting
Troubleshooting Using Service Function
4. T
ro
uble
s
hoo
tin
g
Figure 4-16
A6 Result example
Power On Test Failure Troubleshooting
Table 4-1 represents the contents of the test results and the relationships of failed tests to
probable faulty board assemblies. If the instrument fails the power on test, replace the
faulty board assembly as shown in Table 4-1.
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
Table 4-1
Power on test failure troubleshooting information
Test result
Probable faulty board assembly
A1 F/P
Detect
A2 DC
Source
A3 ADC/
DSP
A4 Local
Synth
A5 Local
Synth
A6 Phase
Detect
A1 Result
###
A2 Result
#
###
A3 Result
#
###
A4 Result
#
###
A5 Result
#
###
A6 Result
#
###
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...