7
169
SCPI Command Reference
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe
7. SCPI Command
Re
fe
re
nc
e
Equivalent key
SP Menu -> Marker Search -> Tracking
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe
Syntax
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe {ON|OFF|1|0}
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:STATe?
Description
Turns on/off markers
Parameter
Equivalent key
SP Menu -> Marker -> Clear Marker Menu -> Marker 1
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:X
Syntax
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:X <numeric>
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:X?
Description
Sets/Reads the marker X value
Parameter
Equivalent key
No equivalent key is available on the front panel.
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:Y
Syntax
:CALCulate:SP[1-1]:TRACe[1-1]:MARKer[1-6]:Y
Description
Reads the marker Y value (Query Only)
TARGet
Set the marker tracking type to 'TARGet'
Description
Description
ON or 1
Set markers to 'ON'
OFF or 0(Preset
value)
Set markers to 'OFF'
<Numeric>
Range
-
Preset value
992.5M
Unit
-
Resolution
-
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...