308
7
SCPI Command Reference
:SENSe:PN[1-1]:DCONverter:FREQuency
:SENSe:PN[1-1]:DCONverter:FREQuency
Syntax
:SENSe:PN[1-1]:DCONverter:FREQuency <numeric>
:SENSe:PN[1-1]:DCONverter:FREQuency?
Description
Sets/Reads the input frequecny to be supplied to the downconverter
Parameter
When the frequency band is 3G to 10GHz
When the frequency band is 9G to 26.5GHz
Equivalent key
PN Menu -> Setup -> Nominal Frequency
:SENSe:PN[1-1]:DCONverter:SSEarch:EXECute
Syntax
:SENSe:PN[1-1]:DCONverter:SSEarch:EXECute
Description
Searches carrier signal and reflects the result to the input frequency of the downconverter
(No query)
Equivalent key
PN Menu -> Setup -> Carrier Search
:SENSe:PN[1-1]:FBANd
Syntax
:SENSe:PN[1-1]:FBANd {BAND1|BAND2|BAND3|BAND4}
:SENSe:PN[1-1]:FBANd?
Description
Sets/Reads frequency band
*1.The softkey is not available when option 011 is installed
<Description>
Range
3G to 10G
Preset value
3G
Unit
Hz
Resolution
100m
<Description>
Range
9G to 26.5G
Preset value
9G
Unit
Hz
Resolution
100m