7
317
SCPI Command Reference
:SENSe:SP[1-1]:FREQuency:STARt
7. SCPI Command
Re
fe
re
nc
e
frequency within the frequency offset, and the conversion mode used in the frequency
offset)
Equivalent key
SP Menu -> Start/Center -> Span
SP Menu -> Stop/Span -> Span
:SENSe:SP[1-1]:FREQuency:STARt
Syntax
:SENSe:SP[1-1]:FREQuency:STARt <numeric>
:SENSe:SP[1-1]:FREQuency:STARt?
Description
Sets/Reads the start value of frequency span
Parameter
When the E5052A is used stand-alone, or with the downconverter turned off
(Allowable setting range varies depending on whether the frequency offset is on or off
, the LO frequency within the frequency offset, and the conversion mode used in the
frequency offset)
When the downconverter is turned on and the RF input is set to 'E5052A Direct' RF
(Allowable setting range varies depending on whether the frequency offset is on or off
, the LO frequency within the frequency offset, and the conversion mode used in the
frequency offset)
<Numeric>
Range
100 to 15M
Preset value
15M
Unit
Hz
Resolution
200m
<Numeric>
Range
9M to 116.99999995G
Preset value
992.5M
Unit
Hz
Resolution
100m
<Description>
Range
9M to 112.99999995G
Preset value
2.9925G
Unit
Hz
Resolution
100m
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...