7
337
SCPI Command Reference
:SOURce:VOLTage:CONTrol:AFC:LIMit:HIGH
7. SCPI Command
Re
fe
re
nc
e
Description
Sets/Reads the maximum number of iterations for the DC control voltage-setting loops
Parameter
Equivalent key
PN Menu -> DC Control Voltage -> Auto Freq Control -> Max Iteration
SP Menu -> DC Control Voltage -> Auto Freq Control -> Max Iteration
FP Menu -> DC Control Voltage -> Auto Freq Control -> Max Iteration
TR Menu -> DC Control Voltage -> Auto Freq Control -> Max Iteration
USER Menu -> DC Control Voltage -> Auto Freq Control -> Max Iteration
:SOURce:VOLTage:CONTrol:AFC:LIMit:HIGH
Syntax
:SOURce:VOLTage:CONTrol:AFC:LIMit:HIGH <numeric>
:SOURce:VOLTage:CONTrol:AFC:LIMit:HIGH?
Description
Sets/Reads the maximum DC control voltage limit
Parameter
Equivalent key
PN Menu -> DC Control Voltage -> Auto Freq Control -> Max Ctrl Voltage Limit
SP Menu -> DC Control Voltage -> Auto Freq Control -> Max Ctrl Voltage Limit
FP Menu -> DC Control Voltage -> Auto Freq Control -> Max Ctrl Voltage Limit
TR Menu -> DC Control Voltage -> Auto Freq Control -> Max Ctrl Voltage Limit
USER Menu -> DC Control Voltage -> Auto Freq Control -> Max Ctrl Voltage Limit
:SOURce:VOLTage:CONTrol:AFC:LIMit:LOW
Syntax
:SOURce:VOLTage:CONTrol:AFC:LIMit:LOW <numeric>
<Numeric>
Range
1 to 99
Preset value
10
Unit
-
Resolution
-
<Numeric>
Range
-15 to 35
Preset value
35
Unit
V
Resolution
100u
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...