Appendix B
489
Status Reporting System
Status Register Structure
B. S
tat
us Re
po
rt
ing
Sy
ste
m
event register.
command will clear all of the bits in the questionable limit TR1 status
event register.
Table B-15
Status Bit Definitions of Questionable Limit TR1 Status Event Register
Bit
Position
Name
Description
0
Not used
Always 0.
1
Trace 1 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 1.
2
Trace 2 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 2.
3
Trace 3 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 3.
4
Trace 4 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 4.
5-15
Not used
Always 0.
Table B-16
Status Bit Definitions of Questionable Limit USER1 Status Event Register
Bit
Position
Name
Description
0
Not used
Always 0.
1
Trace 1 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 1.
2
Trace 2 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 2.
3
Trace 3 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 3.
4
Trace 4 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 4.
5
Trace 5 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 5.
6
Trace 6 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 6.
7
Trace 7 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 7.
8
Trace 8 Limit Test Fail
Set to “0” when a measurement cycle begins;
set to “1” when the measurement cycle finishes and
returns “fail” as the limit test result for trace 8.
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...