7
311
SCPI Command Reference
:SENSe:PN[1-1]:LOBandwidth
7. SCPI Command
Re
fe
re
nc
e
Parameter
Equivalent key
PN Menu -> Setup -> IF Gain
*1
:SENSe:PN[1-1]:LOBandwidth
Syntax
:SENSe:PN[1-1]:LOBandwidth {NARRow|WIDE}
:SENSe:PN[1-1]:LOBandwidth?
Description
Sets/readst phase noise Local bandwidth optimization.
Parameter
Equivalent key
PN Menu -> Setup -> LO PhNoise Optimize
:SENSe:PN[1-1]:SEGTable[:MEASurement][:QUALity]
Syntax
:SENSe:PN[1-1]:SEGTable[:MEASurement][:QUALity] {NORMal|FAST}
:SENSe:PN[1-1]:SEGTable[:MEASurement][:QUALity]?
Description
Sets/Reads the quality level
Parameter
<Numeric>
Range
0 to 50 (standard)
10 (option 011,fixed value)
Preset Value
20
10 (option 011,fixed value)
Unit
dB
Resolution
10
*1.The value of IF Gain cannot be changed when option 011 is installed.
Description
NARRow
Set phase noise Local bandwidth optimization to 'NARRow'
WIDE(Preset
value)
Set phase noise Local bandwidth optimization to 'WIDE'
Description
NORMal(Preset
value)
Set the quality level to 'Normal'