182
Chapter 6
Replacement Procedure
A27 Mass storage Disk Drive Assembly Replacement
Adjustments for the New HDD
Step 1.
Run the adjustment program.
Step 2.
Input the GPIB address of the E5052A.
Step 3.
Input the serial number of the E5052A.
Step 4.
Perform the adjustment using “HDD” in
Spot Adjustment
of the program.
Calibration of the Touch Screen
When you have replaced the HDD on a E5052A, you have to calibrate the touch screen.
Follow the procedure described below to calibrate the touch screen.
Step 1.
Press [System].
Step 2.
Press
Service Menu
.
Step 3.
Press
Test Menu
.
Step 4.
Press
Adjust Touch Screen
.
The touch screen calibration screen (Figure 6-4) appears.
Figure 6-4
Touch Panel Calibration Screen
Step 5.
Touch the x mark on the upper left with your finger. The mark x appears also on the lower
left, upper right, and lower right. Touch the x marks in that order with your finger.
Touching the four locations described above with your finger automatically concludes the
touch screen calibration.
NOTE
With no operation on the touch screen calibration screen for a preset time, it automatically
closes and the previous measurement screen reappears.
Summary of Contents for E5052A
Page 13: ...Contents 13 M 271 N 271 O 271 P 271 R 272 S 273 T 273 U 273 ...
Page 14: ...14 Contents ...
Page 22: ...22 Chapter1 General Information Required Equipment ...
Page 46: ...46 Chapter2 Performance Tests Individual Test Description Step 17 Turn off the power of AMP2 ...
Page 82: ...82 Chapter3 Adjustment Required Adjustment after Replacing Assembly ...
Page 218: ...216 Chapter6 Replacement Procedure A51 LCD Replacement Figure 6 27 A51 LCD Removal ...
Page 222: ...220 Chapter6 Replacement Procedure BNC Connector front Replacement ...
Page 244: ...242 AppendixB System Recovery Calibration of the Touch Screen ...
Page 250: ...248 AppendixC Firmware Update Firmware Update ...
Page 261: ...259 E Power Requirement E Power Requirement ...